Description
Method:
Soluble silica reacts with the molybdate ion in an acid medium to form a green-yellow colored silico-molybdic acid complex that in its turn is converted to a blue complex with the addition of the reducing reagent. Oxalic acid is added to minimize the phosphate interference. The absorbance intensity is proportional to the silica concentration in the sample and is determined at 850 nm.
Features:
Measured parameter | SiO2 | ||||||
Range
|
|||||||
Reproducibility*
|
|||||||
Limit of detection | 0.5 ppb |
Wavelength | 850 nm | ||||
Demineralized water | Not required | ||||
Number of reagents | 4 | ||||
Min. analysis time | 10 minutes | ||||
Default analysis time | 15 minutes | ||||
Reagent consumption**
|
Technical Specifications
Model | 3S-CL |
Sampling Mode | Batch, with freely settable frequency. |
Sample | Pressure: Atmospheric Temperature: 5 ‐ 45°C (41 ‐ 113°F) Flow Rate: 80 to 500 mL/min Connection: 6 mm (¼-in.) |
Drain | Pressure-free drain Connection: 12 mm (½-in.) |
Input Streams | 2 |
Material | Epoxy-coated stainless steel AISI 304 |
Dimensions (H x W x D) | 604 x 380 x 210 mm (23.6 x 14.8 x 8.2 in) |
Weight | 20 kg (44 lbs) |
Power Supply | Input Voltage: 115 VAC,230 VAC Power consumption: max. 80 W |
Output signals | 2 x 4-20 mA analog outputs Modbus via RS485,Ethernet |
Alarms | 2 SPDT programmable voltage-free relays |
Digital Input | None, Online, Start Extra, Skip Idle, Emergency Stop |
Ambient Temperature | 5 – 45 °C (41 – 113 °F) |
Ambient Humidity | 10 – 90 % RH (non-condensing) |
Protection Grade | IP54 (indoor only, outdoor use possible with external cabinet, not included) |